Putting the Instrument into Operation    Related Topics

This section describes the basic steps to be taken when setting up the analyzer for the first time.

  1. Unpacking the instrument

  2. Setting up the instrument on a benchtop or in a rack

  3. Connecting to the AC power supply

  4. Switching on

 

Attention!

Before turning on the instrument, please make sure that the following conditions are fulfilled:

  • Instrument covers are in place and all fasteners are tightened.

  • Fan openings are free from obstructions.

  • Signal levels at the input connectors are all within the specified maxima.

  • Signal outputs are correctly connected and not overloaded.

  • The instrument is dry and shows no condensation.  

Non-observance may cause damage to the instrument!

 

 


Unpacking and Checking the Instrument

To remove the instrument from its packaging and check the equipment for completeness proceed as follows:

  1. Pull off the polyethylene protection pads from the instrument's rear feet and then carefully remove the pads from the instrument handles at the front.

  2. Pull off the corrugated cardboard cover that protects the rear of the instrument.

  3. Carefully unthread the corrugated cardboard cover at the front that protects the instrument handles and remove it.

  4. Check the equipment for completeness using the delivery note and the accessory lists for the various items.

  5. Check the instrument for any damage. If there is damage, immediately contact the carrier who delivered the instrument.

Retain the original packing material. If the instrument needs to be transported or shipped at a later date, you can use the material to prevent control elements and connectors from being damaged.


Instrument Setup

The network analyzer is designed for use under laboratory conditions, either on a bench top or in a rack. The general ambient conditions required at the operating site are as follows:

 

Attention!

To avoid damage of electronic components of the DUT and the analyzer, the operating site must be protected against electrostatic discharge (ESD). ESD is most likely to occur when you connect or disconnect a DUT or test fixture to the analyzer's test ports.

To prevent ESD damage use the wrist strap and grounding cord supplied with the instrument and connect yourself to the GND connector at the front panel.  

 

 

 


Bench Top Operation

If the analyzer is operated on a bench top, the surface should be flat.

The instrument can be used in horizontal position, standing on its feet, or with the support feet on the bottom extended.

 

Caution!

The feet must be fully folded in or out. Only in this way can the stability of the analyzer be guaranteed and reliable operation be ensured.

The total weight on the extended feet (including the analyzer's weight and other units placed on top) must not exceed 500 N. The units must be secured against slipping (e.g. by locking the feet at the upper front frame).

When moving the unit with expanded feet, the feet might collapse and fold in. To avoid injuries, the unit must therefore not be moved with the feet out.

 

 


Mounting in a 19" Rack

Using the adapter R&S ZZA-611 (order number 1096.3302.00, for large size units) or R&S ZZA-511 (order number 1096.3290.00, for compact size units) the instrument can be mounted in 19" racks according to the mounting instructions supplied with the rack adapter.

Attention!

  • Allow for sufficient air supply in the rack.

  • Make sure that there is sufficient space between the ventilation holes and the rack casing.


EMI Suppression

To suppress generated Electromagnetic Interference (EMI), operate the instrument only while it is closed, with all shielding covers fitted. Note the EMC classification in the data sheet.

Use appropriate shielded cables to ensure successful control of electromagnetic radiation during operation, especially for the following connector types:

The use of external accessories for the network analyzers may introduce additional connector, cable, and cable length requirements. Refer to the relevant documentation.


Connecting the Instrument to the AC Supply

The network analyzer is automatically adapted to the AC supply voltage supplied. The supply voltage must be in the range 100 V to 240 V; 50 Hz to 60 Hz. The mains connector is located at the bottom left corner of the rear panel.

The maximum power consumption of the analyzer is 450 W. The typical power consumption of the individual analyzer models are is listed in the "Specifications".  

The network analyzer is protected by two fuses as specified on the label on the power supply. The fuses are located on an AC Fuse Board (order no. 1145.3906.02) which must be replaced to change the fuses. Replacing the AC Fuse Board requires opening the instrument and is described in the service manual.


Power on and off

The mains connector is located at the bottom left corner of the rear panel.

After power-on, the analyzer is in standby or ready state, depending on the state of the STANDBY toggle switch at the front panel when the instrument was switched off for the last time.

The AC power switch can be permanently on. Switching off is required only if the instrument must be completely removed from the AC power supply.


Standby and Ready State

The STANDBY toggle switch is located in the bottom left corner of the front panel.

Caution!

The instrument is still power-supplied while it is in standby mode.

Remote control:

SYSTem:SHUTdown


Replacing Fuses

The DC inputs PORT BIAS at the rear panel are each protected by a fuse IEC 127 - F 250 L (250 mA quick acting).