[SENSe<Chn>:]PAE...    Related Topics

This subsystem configures the measurement of the Power Added Efficiency (PAE).

 

Command tree: [SENSe<Chn>:]PAE

:C

:EXPRession

:K


[SENSe<Chn>:]PAE:C <numeric_value>

Defines the constant c for the PAE measurement.

<Chn>

Channel number used to identify the active trace.

 

<numeric_value>

Value of the constant c in the PAE test model ([SENSe<Chn>:]PAE:EXPRession). The value is used if one of the test models C10 | C1 | CK11 is selected.

Range [def. unit]

-3.4028234664E+038 W/V to +3.4028234664E+038 W/V [W/V].

*RST value

0 W/V

 

SCPI, Command Types

Device-specific, command or query

 

Example:

*RST; CALC:PAR:MEAS "TRC1", "PAE21"

Select the PAE as a measurement parameter for the default trace Trc1. Select port 1 as the DUT input port, port 2 as the DUT output port.

PAE:EXPR CK11

Select the test model involving both constants and no power approximation.

PAE:C 10; K 1  

Set the constant c = UDC / R to 10 W/V and the constant k = 1 W/V2. If the test setup is in accordance with the settings made, the trace shows the PAE for a constant DC supply voltage UDC.

 

 


[SENSe<Chn>:]PAE:EXPRession
C10 | C1 | K101 | CK11

Selects the test model for the PAE measurement.

<Chn>

Channel number used to identify the active trace.

 

C10

The DC power is equal to the constant c ([SENSe<Chn>:]PAE:C) times the DC voltage measured at the DC input DC MEAS –10 V...+10 V.

C1

The DC power is equal to the constant c times the DC voltage measured at the DC input DC MEAS –1 V...+1 V.

K101

The DC power is equal to the constant k ([SENSe<Chn>:]PAE:K) times the DC voltage measured at the DC input DC MEAS –10 V...+10 V times the DC voltage measured at the DC input DC MEAS –1 V...+1 V.

CK11

The DC power is equal to the constant c times the DC voltage measured at the DC input DC MEAS –1 V...+1 V minus k times the square of the DC voltage measured at the DC input DC MEAS –1 V...+1 V.

*RST value

CDC10

 

SCPI, Command Types

Device-specific, command or query

 

Example:

See [SENSe<Chn>:]PAE:C

 


[SENSe<Chn>:]PAE:K <numeric_value>

Defines the constant k for the PAE measurement.

<Chn>

Channel number used to identify the active trace.

 

<numeric_value>

Value of the constant k in the PAE test model ([SENSe<Chn>:]PAE:EXPRession). The value is used if one of the test models K101 | CK11 is selected.

Range [def. unit]

-3.4028234664E+038 W/V2 to +3.4028234664E+038 W/V2 [W/V2].

*RST value

0 W/V

 

SCPI, Command Types

Device-specific, command or query

 