This subsystem configures the measurement of the Power Added Efficiency (PAE).
Command tree: [SENSe<Chn>:]PAE
Defines the constant c for the PAE measurement.
<Chn> |
Channel number used to identify the active trace.
|
<numeric_value> |
Value of the constant c in the PAE test model ([SENSe<Chn>:]PAE:EXPRession). The value is used if one of the test models C10 | C1 | CK11 is selected. |
Range [def. unit] |
-3.4028234664E+038 W/V to +3.4028234664E+038 W/V [W/V]. |
*RST value |
0 W/V
|
SCPI, Command Types |
Device-specific, command or query
|
Example: |
*RST; CALC:PAR:MEAS "TRC1", "PAE21" Select the PAE as a measurement parameter for the default trace Trc1. Select port 1 as the DUT input port, port 2 as the DUT output port. PAE:EXPR CK11 Select the test model involving both constants and no power approximation. PAE:C 10; K 1 Set the constant c = UDC / R to 10 W/V and the constant k = 1 W/V2. If the test setup is in accordance with the settings made, the trace shows the PAE for a constant DC supply voltage UDC.
|
Selects the test model for the PAE measurement.
<Chn> |
Channel number used to identify the active trace.
|
C10 |
The DC power is equal to the constant c ([SENSe<Chn>:]PAE:C) times the DC voltage measured at the DC input DC MEAS –10 V...+10 V. |
C1 |
The DC power is equal to the constant c times the DC voltage measured at the DC input DC MEAS –1 V...+1 V. |
K101 |
The DC power is equal to the constant k ([SENSe<Chn>:]PAE:K) times the DC voltage measured at the DC input DC MEAS –10 V...+10 V times the DC voltage measured at the DC input DC MEAS –1 V...+1 V. |
CK11 |
The DC power is equal to the constant c times the DC voltage measured at the DC input DC MEAS –1 V...+1 V minus k times the square of the DC voltage measured at the DC input DC MEAS –1 V...+1 V. |
*RST value |
CDC10
|
SCPI, Command Types |
Device-specific, command or query
|
Example: |
|
Defines the constant k for the PAE measurement.
<Chn> |
Channel number used to identify the active trace.
|
<numeric_value> |
Value of the constant k in the PAE test model ([SENSe<Chn>:]PAE:EXPRession). The value is used if one of the test models K101 | CK11 is selected. |
Range [def. unit] |
-3.4028234664E+038 W/V2 to +3.4028234664E+038 W/V2 [W/V2]. |
*RST value |
0 W/V
|
SCPI, Command Types |
Device-specific, command or query
|
Example: |
|