[SENSe<Ch>:]FREQuency...    Related Topics

This subsystem sets frequency-related parameters, especially the measurement and display ranges for the different sweep types, and controls the frequency-converting measurement modes (harmonic distortion, mixer mode, etc.).

The frequency ranges for the different instrument models are listed below:

 

ZVB4

ZVB8

ZVB14

ZVB20

Start, Stop

300 kHz to 4 GHz

300 kHz to 8 GHz

10 MHz to 14 GHz

10 MHz to 20 GHz

Center

> 300 kHz to < 4 GHz

> 300 kHz to < 8 GHz

> 10 MHz to < 14 GHz

> 10 MHz to < 20 GHz

Span

2 Hz to 3.9997 GHz

10 mHz to 7.9997 GHz

10 mHz to 19.99 GHz

10 mHz to 19.99 GHz

 

 

 

Use SYSTem:FREQuency? MINimum | MAXimum. to query the frequency range of your analyzer.

Command tree: [SENSe<Ch>:]FREQuency

:CENTer

:CONVersion

:AWReceiver

[:STATe]

:DEVice

:MODE

:NAME

:PCOefficient<Port>

:DEFault

:HARMonic (commands for harmonic distortion measurement)

:ORDer

:RELative

:RPORt

:SPORt

:MIXer...(commands for mixer measurement)

:CW|FIXed

:IMODulation (commands for intermodulation measurement)

:LPNoise

:MODE

:OFFSet

:PWAVes

:WAVes

:SBANd

:SPAN

:STARt

:STOP

 

Note: For a frequency sweep the range can be defined alternatively by a combination of Start/Stop frequencies or a Center frequency and Span. In fixed frequency mode (power sweep or Time/CW Mode; see [SENSe<Chn>:]FUNCtion[:ON]) the measurement frequency is set by means of [SENSe<Ch>:]FREQuency:CW|FIXed.


[SENSe<Ch>:]FREQuency:CENTer <center_frequency>

Defines the center of the measurement and display range for a frequency sweep.

<Ch>

Channel number. If unspecified the numeric suffix is set to 1.

 

<center_frequency>

Center frequency of the sweep.

Range [def. unit]

Depending on the instrument model [Hz]. The increment (parameters UP or DOWN) is 0.1 kHz.

*RST value

Center of the analyzer's maximum frequency range: (fMIN + fMAX)/2

 

SCPI, Command Types

Confirmed, command or query.

 

Example:

FUNC "XFR:POW:RAT  B1, A2"

Activate a frequency sweep and select the ratio B1/A2 as measured parameter for channel and trace no. 1.

SYST:FREQ? MIN; :SYST:FREQ? MAX

Query the frequency range of the analyzer.

FREQ:CENT 100MHz

Set center frequency to 100 MHz.

FREQ:SPAN 50000

Set frequency span to 50 kHz.

 

 

The measurement range defined by means of the center frequency and the current span ([SENSe<Ch>:]FREQuency:SPAN) must not exceed the allowed frequency range of the analyzer. If necessary, the span is reduced to min (Center – fMIN, fMAX – Center).

[SENSe<Ch>:]FREQuency:CONVersion
FUNDamental | HARMonic | SHARmonic | THARmonic | MIXer

Enables the measurement of a harmonic or the fundamental signal for channel <Ch>.

Enabling a harmonics measurement also resets the measured quantity; see program example below.

<Ch>

Channel number.

 

FUNDamental

Measurement of the fundamental signal (no frequency-converting mode).  

HARMonic

Harmonics measurement. The order of the measured harmonic, the source port, and the receive port must be selected using the [SENSe<Ch>:]FREQuency:CONVersion:HARMonic... commands; see example below.

SHARmonic |
THARmonic

Measurement of the 2nd or 3rd harmonic with source port 1 and receive port 2. These parameters have been introduced for compatibilty with the ZVR. While they are active, a query returns HARM.

MIXer

Scalar mixer mode. The mixer mode is configured using the [SENSe<Ch>:]FREQuency:CONVersion:MIXer... commands.

*RST value

FUNDamental

 

SCPI, Command Types

Device-specific, command or query. Returned values: FUND or HARM.

 

Example:

*RST; FREQ:STAR 0.5E+9; STOP 1E+9

Reset the analyzer (activating a frequency sweep) and set the sweep range between 1 GHz and 1.1 GHz.

FREQ:CONV:HARM:ORD 4; SPOR 1; RPOR 2

Configure a 4th harmonic measurement with source port 1 and receive port 2. The measurement is still not enabled.

FREQ:CONV HARM

Enable the configured harmonic measurement.

CALC:PAR:MEAS? "TRC1"

Query the measured quantity. The analyzer measures the wave quantity b2, the response is 'B2D1SAM'.  

FREQ:CONV:HARM:RELative ON

Enable the relative harmonic measurement, creating a new channel and a new trace Trc2.

CALC2:PAR:MEAS? "TRC2"

Query the measured quantity in the new channel no. 2. The analyzer still measures the wave quantity b2, the response is 'B2D1SAM'.   

CALC2:MATH:STAT?

Check whether the active trace in channel no. 2 (i.e. Trc2) is a mathematical trace: the response is 1.   

 

 

 

 


[SENSe<Ch>:]FREQuency:CONVersion:AWReceiver[:STATe] <Boolean>

Qualifies whether a-waves (outgoing waves) are measured at the source or at the receiver frequency. The setting is relevant for all frequency-converting measurements.

<Ch>

Channel number.

 

<Boolean>

OFF - Measurement at source frequency
ON - Measurement at receiver frequency

*RST value

OFF

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

FREQ:CONV:AWR ON

Measure a waves at the receiver frequency.  

 


[SENSe<Ch>:]FREQuency:CONVersion:DEVice:MODE RILI | RILE

Selects the test setup (internal or external sources) for the frequency converter measurement (with option ZVA-K8, Converter Control).

<Ch>

Channel number. This suffix is ignored because the device mode is a global setting.

 

Parameters

RILI - Measurement at source frequency
RILE - Measurement at receiver frequency

*RST value

RILI

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

FREQ:CONV:DEV:NAME 'ZVA-Z110'

Select frequency converter type R&S ZVA-Z110.

SYST:COMM:RDEV:GEN1:DEF 'Ext. Gen. 1', 'SME02', 'gpib0',  '21'

Configure an R&S SME02 generator as external generator no. 1, assigning the name Ext. Gen. 1 and a GPIB address 21.

FREQ:CONV:DEV:MODE RILE

Measure a waves at the receiver frequency.  

 


[SENSe<Ch>:]FREQuency:CONVersion:DEVice:NAME '<Converter Type>'

Selects the frequency converter type for enhanced frequency-converting measurements (with option ZVA-K8, Converter Control).

<Ch>

Channel number. This suffix is ignored because the converter type is a global setting.

 

'<Converter Type>'

'ZVA-Z110' - Frequency converter R&S ZVA-Z110
'' - No frequency converter selected

*RST value

OFF

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

See [SENSe<Ch>:]FREQuency:CONVersion:DEVice:MODE

 

 


[SENSe<Ch>:]FREQuency:CONVersion:DEVice:PCOefficient<Port> <C0>, <C1>, <C2>, <C3>

Defines power coefficients for frequency converters R&S ZVA-ZxxxE.  

<Ch>

Channel number. This suffix is ignored because converter settings are global.

 

<Port>

NWA port number

 

<C0>, <C1>, <C2>, C3>

Coefficient of order 0, 1, 2, and 3.

Range [def. unit]

-500 to +500 [–].

*RST value

Depending on the individual converter

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

See [SENSe<Ch>:]FREQuency:CONVersion:DEVice:PCOefficient:DEFault

 

 


[SENSe<Ch>:]FREQuency:CONVersion:DEVice:PCOefficient<Port>:DEFault <Boolean>

Enables or disables default power coefficients for frequency converters R&S ZVA-ZxxxE.

<Ch>

Channel number. This suffix is ignored because converter settings are global.

 

<Port>

NWA port number. This suffix is ignored because the command affects all ports.

 

<Boolean>

ON - Use default power coefficients
OFF - Use power coefficients defined via [SENSe<Ch>:]FREQuency:CONVersion:DEVice:PCOefficient<Port>

*RST value

ON

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

FREQ:CONV:DEV:NAME 'ZVA-Z110E'

Select frequency converter type R&S ZVA-Z110E.

FREQ:CONV:DEV:PCO3:DEF OFF

FREQ:CONV:DEV:PCO3 0.5479,0.03426,-0.002866,2.124E-006

Re-define the power coefficients for port 3 according to the connected converter unit.

FREQ:CONV:DEV:PCO3:DEF ON

Use default power coefficients for all ports.  

FREQ:CONV:DEV:PCO3?

Query the default power coefficients for port 3. The analyzer returns the predefined values (generally different from the coefficients entered above).

 


[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:ORDer <order>

Selects the order of the harmonic measured if a harmonic measurement is enabled ([SENSe<Ch>:]FREQuency:CONVersion HARMonic).

<Ch>

Channel number.

 

<order>

Order of the harmonic, integer multiple of the fundamental frequency measured.

Range [def. unit]

2 to 100 [–].

*RST value

2

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

See [SENSe<Ch>:]FREQuency:CONVersion

 

 


[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:RELative <Boolean>

Enables or disables the relative harmonic measurement where the harmonic is divided by the fundamental wave.

With enabled relative measurement the analyzer creates a new channel and a new (mathematical) trace; see program example for [SENSe<Ch>:]FREQuency:CONVersion.

<Ch>

Channel number.

 

<Boolean>

Enable (ON) or disable (OFF) the relative harmonic measurement.

*RST value

OFF

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

See [SENSe<Ch>:]FREQuency:CONVersion

 

 


[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:RPORt <port_no>

Selects the receive port for the harmonic measurement.

<Ch>

Channel number.

 

<port_no>

Integer port number.

Range [def. unit]

Depending on the analyzer model [–].

*RST value

2

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

See [SENSe<Ch>:]FREQuency:CONVersion

 

 


[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:SPORt <port_no> | <ext. gen. no.>

Selects the source port for the harmonic measurement.

<Ch>

Channel number.

 

<port_no>

Positive integer analyzer port number.

<ext. gen. no.>

Negative integer external generator number (the minus sign serves as an identifier for external generators; –1 denotes the external generator no. 1)

Range [def. unit]

Depending on the analyzer model and the number of external generators [–].

*RST value

1

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

See [SENSe<Ch>:]FREQuency:CONVersion