[SENSe<Ch>:]BANDwidth...    Related Topics

This subsystem sets the bandwidth of the IF measurement filter (resolution bandwidth).The forms BANDwidth and BWIDth are equivalent.   

   

Command tree: [SENSe<Ch>:]BANDwidth|BWIDth

[:RESolution]

:SELect


[SENSe<Ch>:]BANDwidth|BWIDth[:RESolution] <bandwidth>

Defines the resolution bandwidth of the analyzer (Meas. Bandwidth).

<Ch>

Channel number. If unspecified the numeric suffix is set to 1.

 

<bandwidth>

Resolution bandwidth

Range [def. unit]

1 Hz to 1 MHz (5 (10) MHz with option R&S ZVA-K17, 10 (30) MHz with option R&S ZVA-K7; see Pulsed Measurements) [Hz]. UP and DOWN increment/decrement the bandwidth in 1-2-5 steps for each decade. The analyzer rounds up any entered value between these steps and rounds down values exceeding the maximum bandwidth.  

*RST value

10 kHz

 

SCPI, Command Types

Confirmed, command or query

 

Example:

BAND 1.1

Set a resolution bandwidth of approx. 1.1 Hz for channel 1.

BAND?

The analyzer returns the rounded bandwidth of 2 Hz.

 

 


[SENSe<Ch>:]BANDwidth|BWIDth[:RESolution]:SELect NORMal | HIGH

Defines the selectivity of the IF filter for an unsegmented sweep. The value is also used for all segments of a segmented sweep, provided that separate selectivity setting is disabled ([SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol OFF).

<Ch>

Channel number.

 

NORMal

Use IF filter with normal selectivity and short settling time.

HIGH

Use IF filter with higher selectivity but larger settling time.

*RST value

NORMal

 

SCPI, Command Types

Device-specific, command or query

 

Example:

See [SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect:CONTrol