[SENSe<Ch>:]FREQuency...    Related Topics

This subsystem sets frequency-related parameters, especially the measurement and display ranges for the different sweep types, and controls the frequency-converting measurement modes (harmonic distortion, mixer mode, etc.).

The frequency ranges for the different instrument models are listed below:

 

 

 

ZVA8

ZVA24

ZVA40

ZVA50

ZVA67

Start, Stop

300 kHz to 8 GHz

10 MHz to 25 GHz
(the specified frequency range is 10 MHz to 24 GHz; see data sheet)

10 MHz to 40 GHz

10 MHz to 50 GHz

10 MHz to 67 GHz

Center

> 300 kHz to < 8 GHz

> 10 MHz to < 25 GHz

> 10 MHz to < 40 GHz

> 10 MHz to < 50 GHz

> 10 MHz to < 67 GHz

Span

10 mHz to 7.9997 GHz

10 mHz to 24.99 GHz

10 mHz to 39.99 GHz

10 mHz to 49.99 GHz

10 mHz to 66.99 GHz

 

Use SYSTem:FREQuency? MINimum | MAXimum. to query the frequency range of your analyzer.

Command tree: [SENSe<Ch>:]FREQuency

:CENTer

:CONVersion

:ARBitrary

:PMETer

:AWReceiver

[:STATe]

:DEVice

:MODE

:NAME

:PCOefficient<Port>

:DEFault

:HARMonic (commands for harmonic distortion measurement)

:ORDer

:RELative

:RPORt

:SPORt

:MIXer...(commands for mixer measurement)

:CW|FIXed

:IMODulation (commands for intermodulation measurement)

:LPNoise

:MODE

:OFFSet

:PWAVes

:WAVes

:SBANd

:SPAN

:STARt

:STOP

 

Note: For a frequency sweep the range can be defined alternatively by a combination of Start/Stop frequencies or a Center frequency and Span. In fixed frequency mode (power sweep or Time/CW Mode; see [SENSe<Chn>:]FUNCtion[:ON]) the measurement frequency is set by means of [SENSe<Ch>:]FREQuency:CW|FIXed.


[SENSe<Ch>:]FREQuency:CENTer <center_frequency>

Defines the center of the measurement and display range for a frequency sweep.

<Ch>

Channel number. If unspecified the numeric suffix is set to 1.

 

<center_frequency>

Center frequency of the sweep.

Range [def. unit]

Depending on the instrument model [Hz]. The increment (parameters UP or DOWN) is 0.1 kHz.

*RST value

Center of the analyzer's maximum frequency range: (fMIN + fMAX)/2

 

SCPI, Command Types

Confirmed, command or query.

 

Example:

FUNC "XFR:POW:RAT  B1, A2"

Activate a frequency sweep and select the ratio B1/A2 as measured parameter for channel and trace no. 1.

SYST:FREQ? MIN; :SYST:FREQ? MAX

Query the frequency range of the analyzer.

FREQ:CENT 100MHz

Set center frequency to 100 MHz.

FREQ:SPAN 50000

Set frequency span to 50 kHz.

 

 

The measurement range defined by means of the center frequency and the current span ([SENSe<Ch>:]FREQuency:SPAN) must not exceed the allowed frequency range of the analyzer. If necessary, the span is reduced to min (Center – fMIN, fMAX – Center).

[SENSe<Ch>:]FREQuency:CONVersion
FUNDamental | HARMonic | SHARmonic | THARmonic | MIXer | VMIXer

Enables the measurement of a harmonic or the fundamental signal for channel <Ch>.

Enabling a harmonics measurement also resets the measured quantity; see program example below.

<Ch>

Channel number.

 

FUNDamental

Measurement of the fundamental signal (no frequency-converting mode).  

HARMonic

Harmonics measurement. The order of the measured harmonic, the source port, and the receive port must be selected using the [SENSe<Ch>:]FREQuency:CONVersion:HARMonic... commands; see example below.

SHARmonic |
THARmonic

Measurement of the 2nd or 3rd harmonic with source port 1 and receive port 2. These parameters have been introduced for compatibilty with the ZVR. While they are active, a query returns HARM.

MIXer |
VMIXer

Scalar or vector mixer mode. Both mixer modes are configured using the [SENSe<Ch>:]FREQuency:CONVersion:MIXer... commands.

*RST value

FUNDamental

 

SCPI, Command Types

Device-specific, command or query. Returned values: FUND or HARM.

 

Example:

*RST; FREQ:STAR 0.5E+9; STOP 1E+9

Reset the analyzer (activating a frequency sweep) and set the sweep range between 1 GHz and 1.1 GHz.

FREQ:CONV:HARM:ORD 4; SPOR 1; RPOR 2

Configure a 4th harmonic measurement with source port 1 and receive port 2. The measurement is still not enabled.

FREQ:CONV HARM

Enable the configured harmonic measurement.

CALC:PAR:MEAS? "TRC1"

Query the measured quantity. The analyzer measures the wave quantity b2, the response is 'B2D1SAM'.  

FREQ:CONV:HARM:RELative ON

Enable the relative harmonic measurement, creating a new channel and a new trace Trc2.

CALC2:PAR:MEAS? "TRC2"

Query the measured quantity in the new channel no. 2. The analyzer still measures the wave quantity b2, the response is 'B2D1SAM'.   

CALC2:MATH:STAT?

Check whether the active trace in channel no. 2 (i.e. Trc2) is a mathematical trace: the response is 1.   

 

 


[SENSe<Ch>:]FREQuency:CONVersion:ARBitrary
<numerator>, <denominator>, <offset>, CW | FIXed | SWEep

Defines the receiver frequency for frequency-converting measurements. The receiver frequency is either a range (for frequency sweeps) or a CW frequency (for power, time and CW Mode sweeps). The receiver frequency is valid for all ports.

The frequency formula is applied even if the analyzer returns an error message, because the frequency is outside the allowed range.

<Ch>

Channel number.

 

<numerator>
<denominator>
<offset>

Parameters of the frequency formula. The receiver frequency fr is calculated according to
fr = <numerator>/<denominator>*fb. + <offset>
where fb represents the sweep range (for frequency sweeps, parameter SWEep). For power, time and CW Mode sweeps (parameters CW or FIXed), fb = 0.  

The conversion formula for fr is ZVR-compatible but differs from manual control where the ratio <numerator>/<denominator> also multiplies the <offset>.

Range [def. unit]

The converted frequency or frequency range must be within the analyzer's frequency range, depending on the instrument model [Hz]. In addition the following restrictions apply:
<numerator> – Values are rounded to positive or negative integer numbers; zero is not allowed [1]
<denominator> – Values are rounded to positive integer numbers [1]
<offset> – Values are rounded to (positive or negative) multiples of 1 Hz [Hz].

*RST value

The default frequency/frequency range corresponds to the sweep range or CW frequency of the analyzer, i.e.
<numerator> = <denominator> = 1
<offset> = 0 Hz

 

Sweep type Parameters

SWEep – A frequency sweep is active, the command defines a frequency range.
CW | FIXed – A power, time or CW Mode sweeps is active, the command defines a fixed frequency.

*RST value

SWEep

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

*RST; FREQ:STAR 1E+9; STOP 1.1E+9

Reset the analyzer (activating a frequency sweep) and set the sweep range between 1 GHz and 1.1 GHz.

SENS:FREQ:CONV:ARB  2, 1, 1E+9, SWE

Convert the receiver frequency to the range between 3 GHz and 3.2 GHz.

 

 


[SENSe<Ch>:]FREQuency:CONVersion:ARBitrary:PMETer<Mtr>
<numerator>, <denominator>, <offset>, CW | FIXed | SWEep

Defines the receiver frequency for frequency-converting measurements. The receiver frequency is either a range (for frequency sweeps) or a CW frequency (for power, time and CW Mode sweeps). The receiver frequency is valid for all ports.

The frequency formula is applied even if the analyzer returns an error message, because the frequency is outside the allowed range.

<Ch>

Channel number.

 

<Mtr>

Power meter number

 

<numerator>
<denominator>
<offset>

Parameters of the frequency formula. The receiver frequency fr is calculated according to
fr = <numerator>/<denominator>*fb. + <offset>
where fb represents the sweep range (for frequency sweeps, parameter SWEep). For power, time and CW Mode sweeps (parameters CW or FIXed), fb = 0.  

The conversion formula for fr is ZVR-compatible but differs from manual control where the ratio <numerator>/<denominator> also multiplies the <offset>.

Range [def. unit]

The converted frequency or frequency range must be within the power meter's receive frequency range [Hz]. In addition the following restrictions apply:
<numerator> – Values are rounded to positive or negative integer numbers; zero is not allowed [1]
<denominator> – Values are rounded to positive integer numbers [1]
<offset> – Values are rounded to (positive or negative) multiples of 1 Hz [Hz].

*RST value

The default frequency/frequency range corresponds to the sweep range or CW frequency of the analyzer, i.e.
<numerator> = <denominator> = 1
<offset> = 0 Hz

 

Sweep type Parameters

SWEep – A frequency sweep is active, the command defines a frequency range.
CW | FIXed – A power, time or CW Mode sweeps is active, the command defines a fixed frequency.

*RST value

SWEep

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

*RST; FREQ:STAR 1E+9; STOP 1.1E+9

Reset the analyzer (activating a frequency sweep) and set the sweep range between 1 GHz and 1.1 GHz.

SYST:COMM:RDEV:PMET1:DEF 'USB Power Meter', 'NRP-Z55', 'usb', '100045'

Configure an R&S NRP power meter as external power meter no. 1, assigning the name USB Power Meter and an serial number 100045.

SENS:FREQ:CONV:ARB:PMET  2, 1, 1E+9, SWE

Convert the receiver frequency to the range between 3 GHz and 3.2 GHz.

 

 


[SENSe<Ch>:]FREQuency:CONVersion:AWReceiver[:STATe] <Boolean>

Qualifies whether a-waves (outgoing waves) are measured at the source or at the receiver frequency. The setting is relevant for all frequency-converting measurements.

<Ch>

Channel number.

 

<Boolean>

OFF - Measurement at source frequency
ON - Measurement at receiver frequency

*RST value

OFF

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

FREQ:CONV:AWR ON

Measure a waves at the receiver frequency.  

 


[SENSe<Ch>:]FREQuency:CONVersion:DEVice:MODE RILI | RILE

Selects the test setup (internal or external sources) for the frequency converter measurement (with option ZVA-K8, Converter Control).

<Ch>

Channel number. This suffix is ignored because the device mode is a global setting.

 

Parameters

RILI - Measurement at source frequency
RILE - Measurement at receiver frequency

*RST value

RILI

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

FREQ:CONV:DEV:NAME 'ZVA-Z110'

Select frequency converter type R&S ZVA-Z110.

SYST:COMM:RDEV:GEN1:DEF 'Ext. Gen. 1', 'SME02', 'gpib0',  '21'

Configure an R&S SME02 generator as external generator no. 1, assigning the name Ext. Gen. 1 and a GPIB address 21.

FREQ:CONV:DEV:MODE RILE

Measure a waves at the receiver frequency.  

 


[SENSe<Ch>:]FREQuency:CONVersion:DEVice:NAME '<Converter Type>'

Selects the frequency converter type for enhanced frequency-converting measurements (with option ZVA-K8, Converter Control).

<Ch>

Channel number. This suffix is ignored because the converter type is a global setting.

 

'<Converter Type>'

'ZVA-Z110' - Frequency converter R&S ZVA-Z110
'' - No frequency converter selected

*RST value

OFF

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

See [SENSe<Ch>:]FREQuency:CONVersion:DEVice:MODE

 

 


[SENSe<Ch>:]FREQuency:CONVersion:DEVice:PCOefficient<Port> <C0>, <C1>, <C2>, <C3>

Defines power coefficients for frequency converters R&S ZVA-ZxxxE.  

<Ch>

Channel number. This suffix is ignored because converter settings are global.

 

<Port>

NWA port number

 

<C0>, <C1>, <C2>, C3>

Coefficient of order 0, 1, 2, and 3.

Range [def. unit]

-500 to +500 [–].

*RST value

Depending on the individual converter

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

See [SENSe<Ch>:]FREQuency:CONVersion:DEVice:PCOefficient:DEFault

 

 


[SENSe<Ch>:]FREQuency:CONVersion:DEVice:PCOefficient<Port>:DEFault <Boolean>

Enables or disables default power coefficients for frequency converters R&S ZVA-ZxxxE.

<Ch>

Channel number. This suffix is ignored because converter settings are global.

 

<Port>

NWA port number. This suffix is ignored because the command affects all ports.

 

<Boolean>

ON - Use default power coefficients
OFF - Use power coefficients defined via [SENSe<Ch>:]FREQuency:CONVersion:DEVice:PCOefficient<Port>

*RST value

ON

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

FREQ:CONV:DEV:NAME 'ZVA-Z110E'

Select frequency converter type R&S ZVA-Z110E.

FREQ:CONV:DEV:PCO3:DEF OFF

FREQ:CONV:DEV:PCO3 0.5479,0.03426,-0.002866,2.124E-006

Re-define the power coefficients for port 3 according to the connected converter unit.

FREQ:CONV:DEV:PCO3:DEF ON

Use default power coefficients for all ports.  

FREQ:CONV:DEV:PCO3?

Query the default power coefficients for port 3. The analyzer returns the predefined values (generally different from the coefficients entered above).

 


[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:ORDer <order>

Selects the order of the harmonic measured if a harmonic measurement is enabled ([SENSe<Ch>:]FREQuency:CONVersion HARMonic).

<Ch>

Channel number.

 

<order>

Order of the harmonic, integer multiple of the fundamental frequency measured.

Range [def. unit]

2 to 100 [–].

*RST value

2

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

See [SENSe<Ch>:]FREQuency:CONVersion

 

 


[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:RELative <Boolean>

Enables or disables the relative harmonic measurement where the harmonic is divided by the fundamental wave.

With enabled relative measurement the analyzer creates a new channel and a new (mathematical) trace; see program example for [SENSe<Ch>:]FREQuency:CONVersion.

<Ch>

Channel number.

 

<Boolean>

Enable (ON) or disable (OFF) the relative harmonic measurement.

*RST value

OFF

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

See [SENSe<Ch>:]FREQuency:CONVersion

 

 


[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:RPORt <port_no>

Selects the receive port for the harmonic measurement.

<Ch>

Channel number.

 

<port_no>

Integer port number.

Range [def. unit]

Depending on the analyzer model [–].

*RST value

2

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

See [SENSe<Ch>:]FREQuency:CONVersion

 

 


[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:SPORt <port_no> | <ext. gen. no.>

Selects the source port for the harmonic measurement.

<Ch>

Channel number.

 

<port_no>

Positive integer analyzer port number.

<ext. gen. no.>

Negative integer external generator number (the minus sign serves as an identifier for external generators; –1 denotes the external generator no. 1)

Range [def. unit]

Depending on the analyzer model and the number of external generators [–].

*RST value

1

 

SCPI, Command Types

Device-specific, command or query.

 

Example:

See [SENSe<Ch>:]FREQuency:CONVersion