Start Cal    Related Topics

The Start Cal submenu selects the calibrated ports and the calibration type.

The calibration types depend on the number of test ports of the analyzer. For a four-port unit:

Use Other... if you wish to perform several calibrations in a single measurement sequence and with a minimum of effort. The Other... calibration wizard allows you to measure a particular standard and reuse the results for different calibration types.  

One-port to 4-port calibrations can be performed automatically using the Calibration Unit (accessory R&S ZV-Z51/-Z52...). For calibrations with more than 4 ports use the calibration units R&S ZV-Z58/-Z59.

Performing a manual calibration

Performing an automatic calibration

Remote control:

[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
'
<cal_name>', REFL | RSHort | FOPort | FRTRans | OPTPort | TOSM | TOM | TRM | TRL | TNA | UOSM, <port_no>[,<port_no>][,<port_no>][,<port_no>]


One Port P1, One Port P2

Opens a submenu to select a one-port calibration at test ports PORT 1 or PORT 2. The two submenus are identical:

One-port calibrations can be performed automatically or manually.

The following menu commands calls up the start dialog of the calibration wizard to start a manual calibration:

Remote control:

[SENSe<Ch>:]CORRection:COLLect:AUTO ' ', <port_no>  (for matching analyzer and cal unit ports)
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
 (for arbitrary pairs of ports)

[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
"
<cal_name>", REFLshort | FOPort , <port_no>


Two Port P1 P2

Opens a submenu to select a two-port calibration at test ports PORT 1 and PORT 2:

Two-port calibrations can be performed automatically or manually.

The following menu commands calls up the start dialog of the calibration wizard to start a manual calibration:

Use Start Cal – Other... to select a UTOM calibration.  

Remote control:

[SENSe<Ch>:]CORRection:COLLect:AUTO ' ', <port_no>  (for matching analyzer and cal unit ports)
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
 (for arbitrary pairs of ports)

[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
"
<cal_name>", FRTRans | OPTPort | TOSM | TOM | TRL | TNA, <port_no>,<port_no>


Opens a submenu to select a three-port TOSM calibration at test ports PORT 1, PORT 2 and PORT 3:

Three-port calibrations can be performed automatically or manually.

Use Start Cal – Other... to select other 3-port calibration types.  

Remote control:

[SENSe<Ch>:]CORRection:COLLect:AUTO ' ', <port_no>  (for matching analyzer and cal unit ports)
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
 (for arbitrary pairs of ports)

[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
"
<cal_name>", TOSM, <port_no>, <port_no>, <port_no>


Opens a submenu to select a four-port TOSM calibration at test ports PORT 1 to PORT 4:

Four-port calibrations can be performed automatically or manually.

Use Start Cal – Other... to select other 4-port calibration types.  

Remote control:

[SENSe<Ch>:]CORRection:COLLect:AUTO ' ', <port_no>  (for matching analyzer and cal unit ports)
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs
 (for arbitrary pairs of ports)

[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
"
<cal_name>", TOSM, <port_no>, <port_no>, <port_no>, <port_no>