CALCulate<Ch>:TRANsform:VNETworks...    Related Topics

This subsystem defines the circuit models for single ended and balanced port (de-)embedding and activates the (de-)embedding function.

 

 

Command tree: CALCulate<Ch>:TRANsform:VNETworks

:BALanced

:DEEMbedding<Log_pt>

:PARameters

:C<nr>

:L<nr>

:R<nr>

[:STATe]

:TNDefinition

:EMBedding<Log_pt>

:PARameters

:C<nr>

:L<nr>

:R<nr>

[:STATe]

:TNDefinition

:GLOop

:DEEMbedding

:PARameters

:C<nr>

:L<nr>

:R<nr>

[:STATe]

:TNDefinition

:EMBedding

:PARameters

:C<nr>

:L<nr>

:R<nr>

[:STATe]

:TNDefinition

:PPAir

:DEEMbedding

:DEFine

:DELete

:PARameters

:C<nr>

:L<nr>

:R<nr>

[:STATe]

:TNDefinition

:EMBedding

:DEFine

:DELete

:PARameters

:C<nr>

:L<nr>

:R<nr>

[:STATe]

:TNDefinition

:SENDed

:DEEMbedding<Ph_pt>

:PARameters

:C<nr>

:L<nr>

:R<nr>

[:STATe]

:TNDefinition

:EMBedding<Ph_pt>

:PARameters

:C<nr>

:L<nr>

:R<nr>

[:STATe]

:TNDefinition

The circuit models are referenced by means of predefined character data parameters. They are different for single ended port, balanced port, and ground loop (de-)embedding.

 

 

Circuit models for balanced port and port pair embedding/deembedding:

Parameter

Circuit model

Capacitances

Resistances

Inductances

FIMPort

File import, no circuit model

STSL

Serial Touchstone (.s2p) data, shunt L

R1

L1

STSC

Serial Touchstone (.s2p) data, shunt C

C1

R1

SLST

Shunt L, serial Touchstone (.s2p) data

R1

L1

SCST

Shunt C, serial Touchstone (.s2p) data

C1

R1

CSSL

Serial Cs, shunt L

C1, C2

R1, R2, R3

L1

LSSC

Serial Ls, shunt C

C1

R1, R2, R3

L1, L2

CSSC

Serial Cs, shunt C

C1, C2, C3

R1, R2, R3

LSSL

Serial Ls, shunt L

R1, R2, R3

L1, L2, L3

SLCS

Shunt L, serial Cs

C1, C2

R1, R2, R3

L1

SCLS

Shunt C, serial Ls

C1

R1, R2, R3

L1, L2

SCCS

Shunt C, serial Cs

C1, C2, C3

R1, R2, R3

SLLS

Shunt L, serial Ls

R1, R2, R3

L1, L2, L3

 

 

 

 

Circuit models for single ended port embedding/deembedding:

Parameter

Circuit model

Capacitances

Resistances

Inductances

FIMPort

File import, no circuit model

CSL

Serial C, shunt L

C1

R1, R2

L1

LSC

Serial L, shunt C

C1

R1, R2

L1

CSC

Serial C, shunt C

C1, C2

R1, R2

LSL

Serial L, shunt L

R1, R2

L1, L2

SLC

Shunt L, serial C

C1

R1, R2

L1

SCL

Shunt C, serial L

C1

R1, R2

L1

SCC

Shunt C, serial C

C1, C2

R1, R2

SLL

Shunt L, serial L

R1, R2

L1, L2

 

 

 

 

Circuit models for ground loop embedding/deembedding:

Parameter

Circuit model

Capacitances

Resistances

Inductances

FIMPort

File import, no circuit model

SL

Shunt L

R

L

SC

Shunt C

C

R

 


CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<Log_pt>:PARameters:C<no>
STSC | SCST | CSSL | LSSC | CSSC | SLCS | SCLS | SCCS,<numeric_value>

Specifies the capacitance values C1, C2, C3 in the different circuit models for balanced port deembedding.  

<Ch>

Channel number. If unspecified the numeric suffix is set to 1.  

 

<Log_pt>

Logical port number (balanced port)

 

<no>

Number of capacitance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of capacitances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Capacitance C<no> for the specified circuit model.

Range [def. unit]

–1mF to 1 mF [F]. The increment is 1 fF (10–15 F).

*RST value

1 pF (10–12 F)

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the capacitance value for the specified circuit model.

 

Example:

*RST; :SOUR:LPOR1 1,2; :CALC:TRAN:VNET:BAL:DEEM:PAR:C2? CSSL

Create a balanced port and query the default capacitance C2 for the Serial Cs, shunt L circuit model. The response is 1E-012 (1 pF).

CALC:TRAN:VNET:BAL:DEEM:PAR:C2 CSSL, 2.2E-12

Increase the capacitance to 2.2 pF.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<Log_pt>:PARameters:L<no>
STSL | SLST | CSSL | LSSC | LSSL | SLCS | SCLS | SLLS,<numeric_value>

Specifies the inductance values L1, L2, L3 in the different circuit models for balanced port deembedding.  

<Ch>

Channel number  

 

<Log_pt>

Logical port number (balanced port)

 

<no>

Number of inductance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of inductances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Inductance L<no> for the specified circuit model.

Range [def. unit]

–1H to 1 H [H]. The increment is 1 pH (10–12 H).

*RST value

1 nH (10–9 H)

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the inductance value for the specified circuit model.

 

Example:

*RST; :SOUR:LPOR1 1,2; :CALC:TRAN:VNET:BAL:DEEM:PAR:L1? CSSL

Create a balanced port and query the default inductance L1 for the Serial Cs, shunt L circuit model. The response is 1E-009 (1 nH).

CALC:TRAN:VNET:BAL:DEEM:PAR:L1 CSSL, 2.2E-9

Increase the inductance to 2.2 nH.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<Log_pt>:PARameters:R<no>
STSL | STSC | SLST | SCST | CSSL | LSSC | CSSC | LSSL | SLCS | SCLS | SCCS | SLLS,<numeric_value>

Specifies the resistance values R1, R2, R3 in the different circuit models for balanced port deembedding.  

<Ch>

Channel number  

 

<Log_pt>

Logical port number (balanced port)

 

<no>

Number of resistance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of resistances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Resistance R<no> for the specified circuit model.

Range [def. unit]

–10 MΩ to 10 MΩ [Ω]. The increment is 1 mΩ (10–3 Ω).

*RST value

0 Ω for all resistances connected in series with an inductance

10 MΩ for all resistances connected in parallel with a capacitance

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the resistance value for the specified circuit model.

 

Example:

*RST; :SOUR:LPOR1 1,2; :CALC:TRAN:VNET:BAL:DEEM:PAR:R1? CSSL; R2? CSSL; R3? CSSL

Create a balanced port and query the default resistances for the Serial Cs, shunt L circuit model. The response is 10000000000;10000000000; 0.

CALC:TRAN:VNET:BAL:DEEM:PAR:R3 CSSL, 2.2E+3

Increase the resistance R3 to 2.2 kΩ.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<Log_pt>[:STATe] <Boolean>

Enables or disables the deembedding function for balanced ports. It is allowed to change the circuit model and its parameters while deembedding is enabled.

<Ch>

Channel number  

 

<Log_pt>

Logical port number (balanced port)

 

<Boolean>

ON – Deembedding active
OFF – Deembedding inactive

*RST value

OFF [–]

 

SCPI, Command Types

Device-specific, command or query

 

Example:

*RST; SOUR:LPOR1 1,2; LPOR2 3,4

Define a balanced port configuration.

CALC:TRAN:VNET:BAL:DEEM:TND CSSL

Select the Serial Cs, shunt L circuit model for deembedding.

CALC:TRAN:VNET:BAL:DEEM:PAR:R3 CSSL, 2.2E+3; :CALC:TRAN:VNET:BAL:DEEM ON

Increase the resistance R3 for the Serial Cs, shunt L circuit model to 2.2 kΩ and enable deembedding.

 

 


CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding<Log_pt>:TNDefinition
FIMPort | STSL | STSC | SLST | SCST | CSSL | LSSC | CSSC | LSSL | SLCS | SCLS | SCCS | SLLS

Selects the circuit model for balanced port deembedding.

<Ch>

Channel number  

 

<Log_pt>

Logical port number (balanced port)

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

CSSL

 

SCPI, Command Types

Device-specific, command or query.  

 

Example:

See CALCulate<Ch>:TRANsform:VNETworks:BALanced:DEEMbedding[:STATe]

 

 


CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<Log_pt>:PARameters:C<no>
STSC | SCST | CSSL | LSSC | CSSC | SLCS | SCLS | SCCS,<numeric_value>

Specifies the capacitance values C1, C2, C3 in the different circuit models for balanced port embedding.  

<Ch>

Channel number

 

<Log_pt>

Logical port number (balanced port)

 

<no>

Number of capacitance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of capacitances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Capacitance C<no> for the specified circuit model.

Range [def. unit]

–1mF to 1 mF [F]. The increment is 1 fF (10–15 F).

*RST value

1 pF (10–12 F)

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the capacitance value for the specified circuit model.

 

Example:

*RST; :SOUR:LPOR1 1,2; :CALC:TRAN:VNET:BAL:EMB:PAR:C2? CSSL

Create a balanced port and query the default capacitance C2 for the Serial Cs, shunt L circuit model. The response is 1E-012 (1 pF).

CALC:TRAN:VNET:BAL:EMB:PAR:C2 CSSL, 2.2E-12

Increase the capacitance to 2.2 pF.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<Log_pt>:PARameters:L<no>
STSL | SLST | CSSL | LSSC | LSSL | SLCS | SCLS | SLLS,<numeric_value>

Specifies the inductance values L1, L2, L3 in the different circuit models for balanced port embedding.  

<Ch>

Channel number  

 

<Log_pt>

Logical port number (balanced port)

 

<no>

Number of inductance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of inductances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Inductance L<no> for the specified circuit model.

Range [def. unit]

–1H to 1 H [H]. The increment is 1 pH (10–12 H).

*RST value

1 nH (10–9 H)

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the inductance value for the specified circuit model.

 

Example:

*RST; :SOUR:LPOR1 1,2; :CALC:TRAN:VNET:BAL:EMB:PAR:L1? CSSL

Create a balanced port and query the default inductance L1 for the Serial Cs, shunt L circuit model. The response is 1E-009 (1 nH).

CALC:TRAN:VNET:BAL:EMB:PAR:L1 CSSL, 2.2E-9

Increase the inductance to 2.2 nH.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<Log_pt>:PARameters:R<no>
STSL | STSC | SLST | SCST | CSSL | LSSC | CSSC | LSSL | SLCS | SCLS | SCCS | SLLS,<numeric_value>

Specifies the resistance values R1, R2, R3 in the different circuit models for balanced port embedding.  

<Ch>

Channel number  

 

<Log_pt>

Logical port number (balanced port)

 

<no>

Number of resistance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of resistances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Resistance R<no> for the specified circuit model.

Range [def. unit]

–10 MΩ to 10 MΩ [Ω]. The increment is 1 mΩ (10–3 Ω).

*RST value

0 Ω for all resistances connected in series with an inductance

10 MΩ for all resistances connected in parallel with a capacitance

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the resistance value for the specified circuit model.

 

Example:

*RST; :SOUR:LPOR1 1,2; :CALC:TRAN:VNET:BAL:EMB:PAR:R1? CSSL; R2? CSSL; R3? CSSL

Create a balanced port and query the default resistances for the Serial Cs, shunt L circuit model. The response is 10000000000;10000000000; 0.

CALC:TRAN:VNET:BAL:EMB:PAR:R3 CSSL, 2.2E+3

Increase the resistance R3 to 2.2 kΩ.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<Log_pt>[:STATe] <Boolean>

Enables or disables the embedding function for balanced ports. It is allowed to change the circuit model and its parameters while embedding is enabled.

<Ch>

Channel number  

 

<Log_pt>

Logical port number (balanced port)

 

<Boolean>

ON – Embedding active
OFF – Embedding inactive

*RST value

OFF [–]

 

SCPI, Command Types

Device-specific, command or query

 

Example:

*RST; SOUR:LPOR1 1,2; LPOR2 3,4

Define a balanced port configuration.

CALC:TRAN:VNET:BAL:EMB:TND CSSL

Select the Serial Cs, shunt L circuit model for embedding.

CALC:TRAN:VNET:BAL:EMB:PAR:R3 CSSL, 2.2E+3; :CALC:TRAN:VNET:BAL:EMB ON

Increase the resistance R3 for the Serial Cs, shunt L circuit model to 2.2 kΩ and enable embedding.

 

 


CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<Log_pt>:TNDefinition
FIMPort | STSL | STSC | SLST | SCST | CSSL | LSSC | CSSC | LSSL | SLCS | SCLS | SCCS | SLLS

Selects the circuit model for balanced port embedding.  

<Ch>

Channel number  

 

<Log_pt>

Logical port number (balanced port)

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

CSSL

 

SCPI, Command Types

Device-specific, command or query.  

 

Example:

See CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMbedding[:STATe]

 

 


CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding:PARameters:C
SC, <capacitance>

Specifies the capacitance value for ground loop deembedding.  

<Ch>

Channel number  

 

Parameters

SC – Shunt C circuit model (character data); see circuit models.

<capacitance>

Capacitance C for ground loop deembedding.

Range [def. unit]

–1mF to 1 mF [F]. The increment is 1 fF (10–15 F).

*RST value

1 pF (10–12 F)

 

SCPI, Command Types

Device-specific, command or query. In the query form, only the first parameter is required; the <capacitance> must be omitted.

 

Example:

*RST; :CALC:TRAN:VNET:GLO:DEEM:PAR:C? SC

Query the default capacitance for ground loop deembedding. The response is 1E-012 (1 pF).

CALC:TRAN:VNET:GLO:DEEM:PAR:C SC, 2.2E-12

Increase the capacitance to 2.2 pF.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding:PARameters:L
SL, <inductance>

Specifies the inductance value for ground loop deembedding.  

<Ch>

Channel number.  

 

Parameters

SL – Shunt L circuit model (character data); see circuit models.

<inductance>

Inductance L for ground loop deembedding.

Range [def. unit]

–1H to 1 H [H]. The increment is 1 pH (10–12 H).

*RST value

1 nH (10–9 H)

 

SCPI, Command Types

Device-specific, command or query. In the query form, only the first parameter is required; the <inductance> must be omitted.

 

Example:

*RST; :CALC:TRAN:VNET:GLO:DEEM:PAR:L? SL

Query the default inductance for ground loop deembedding. The response is 1E-009 (1 nH).

CALC:TRAN:VNET:GLO:DEEM:PAR:L SL, 2.2E-9

Increase the inductance to 2.2 nH.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding:PARameters:R
SC | SL, <resistance>

Specifies the resistance value R in the different circuit models for ground loop deembedding.  

<Ch>

Channel number.  

 

Parameters

Possible circuit models (character data); see circuit models.

<resistance>

Resistance R for the specified circuit model.

Range [def. unit]

–10 MΩ to 10 MΩ [Ω]. The increment is 1 mΩ (10–3 Ω).

*RST value

0 Ω if the resistance is connected in series with an inductance (SL)

10 MΩ if the resistance is connected in parallel with a capacitance (SC))

 

SCPI, Command Types

Device-specific, command or query. In the query form, only the first parameter is required; the <resistance> parameter must be omitted. The command returns the resistance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:GLO:DEEM:PAR:R? SC; R? SL

Query the default resistances for ground loop deembedding. The response is 10000000; 0.

CALC:TRAN:VNET:GLO:DEEM:PAR:R SC, 2.2E+3

Increase the resistance for the Shunt C model  to 2.2 kΩ.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding[:STATe] <Boolean>

Enables or disables the deembedding function for ground loops. It is allowed to change the circuit model and its parameters while deembedding is enabled.

<Ch>

Channel number.  

 

<Boolean>

ON – Deembedding active
OFF – Deembedding inactive

*RST value

OFF [–]

 

SCPI, Command Types

Device-specific, command or query

 

Example:

CALC:TRAN:VNET:GLO:DEEM:TND SL

Select the Shunt L circuit model for deembedding.

CALC:TRAN:VNET:GLO:DEEM:PAR:R SL, 2.2E+3; :CALC:TRAN:VNET:GLO:DEEM ON

Increase the resistance for the Shunt L circuit model to 2.2 kΩ and enable deembedding.

 

 


CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding:TNDefinition
FIMPort | SL | SC

Selects the circuit model for ground loop deembedding.  

<Ch>

Channel number.  

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

FIMPort

 

SCPI, Command Types

Device-specific, command or query.  

 

Example:

See CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding[:STATe]

 

 


CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding:PARameters:C
CSL | LSC | CSC | SLC | SCL | SCC,<capacitance>

Specifies the capacitance values C1, C2 in the different circuit models for ground loop embedding.  

<Ch>

Channel number.  

 

Parameters

SC – Shunt C circuit model (character data); see circuit models.

<capacitance>

Capacitance C for ground loop embedding.

Range [def. unit]

–1mF to 1 mF [F]. The increment is 1 fF (10–15 F).

*RST value

1 pF (10–12 F)

 

SCPI, Command Types

Device-specific, command or query. In the query form, only the first parameter is required; the <capacitance> must be omitted.

 

Example:

*RST; :CALC:TRAN:VNET:GLO:EMB:PAR:C? SC

Query the default capacitance for ground loop embedding. The response is 1E-012 (1 pF).

CALC:TRAN:VNET:GLO:EMB:PAR:C SC, 2.2E-12

Increase the capacitance to 2.2 pF.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding:PARameters:L
SL, <inductance>

Specifies the inductance value for ground loop embedding.  

<Ch>

Channel number.  

 

Parameters

SL – Shunt L circuit model (character data); see circuit models.

<inductance>

Inductance L for ground loop embedding.

Range [def. unit]

–1H to 1 H [H]. The increment is 1 pH (10–12 H).

*RST value

1 nH (10–9 H)

 

SCPI, Command Types

Device-specific, command or query. In the query form, only the first parameter is required; the <inductance> must be omitted.

 

Example:

*RST; :CALC:TRAN:VNET:GLO:EMB:PAR:L? SL

Query the default inductance for ground loop embedding. The response is 1E-009 (1 nH).

CALC:TRAN:VNET:GLO:EMB:PAR:L SL, 2.2E-9

Increase the inductance to 2.2 nH.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding:PARameters:R
SC | SL,<resistance>

Specifies the resistance value R in the different circuit models for ground loop embedding.  

<Ch>

Channel number.  

 

Parameters

Possible circuit models (character data); see circuit models.

<resistance>

Resistance R for the specified circuit model.

Range [def. unit]

–10 MΩ to 10 MΩ [Ω]. The increment is 1 mΩ (10–3 Ω).

*RST value

0 Ω if the resistance is connected in series with an inductance (SL)

10 MΩ if the resistance is connected in parallel with a capacitance (SC)

 

SCPI, Command Types

Device-specific, command or query. In the query form, only the first parameter is required; the <resistance> parameter must be omitted. The command returns the resistance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:GLO:EMB:PAR:R? SC; R? SL

Query the default resistances for ground loop embedding. The response is 10000000; 0.

CALC:TRAN:VNET:GLO:EMB:PAR:R SC, 2.2E+3

Increase the resistance for the Shunt C model  to 2.2 kΩ.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding[:STATe] <Boolean>

Enables or disables the embedding function for ground loops. It is allowed to change the circuit model and its parameters while embedding is enabled.

<Ch>

Channel number.  

 

<Boolean>

ON – Embedding active
OFF – Embedding inactive

*RST value

OFF [–]

 

SCPI, Command Types

Device-specific, command or query

 

Example:

CALC:TRAN:VNET:GLO:EMB:TND SL

Select the Shunt L circuit model for embedding.

CALC:TRAN:VNET:GLO:EMB:PAR:R SL, 2.2E+3; :CALC:TRAN:VNET:GLO:EMB ON

Increase the resistance for the Shunt L circuit model to 2.2 kΩ and enable embedding.

 

 


CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding:TNDefinition
FIMPort | SL | SC

Selects the circuit model for ground loop embedding.  

<Ch>

Channel number.  

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

FIMPort

 

SCPI, Command Types

Device-specific, command or query.  

 

Example:

See CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding[:STATe]

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>:DEFine
<first port no>, <second port no>{, <first port no>, <second port no>}

Defines a list of port pairs for port pair deembedding. The command can be used repeatedly to extend or overwrite the list.  

<Ch>

Channel number   

 

<List_idx>

List index, denotes the current number of the defined pair of ports within the list. If several port pairs are defined, <List_idx> is the number of the first port pair.  

 

<first port no>,
<second port no>

Two different analyzer ports

Range [def. unit]

1 to 4 (depending on analyzer model) [–]

*RST value

 

SCPI, Command Types

Device-specific, no query

 

Example:

See CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>[:STATe]

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>:DELete

Deletes the previously defined list of port pairs for port pair deembedding.  

<Ch>

Channel number   

 

<List_idx>

List index, is available for systematic reasons, but must be always 1 or omitted. The list index in the other ...PPAir:DEEMbedding... commands denotes the current number of a pair of ports within the list.  

 

*RST value

 

SCPI, Command Types

Device-specific, no query

 

Example:

See CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>[:STATe]  

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>:PARameters:C<no>
STSC | SCST | CSSL | LSSC | CSSC | SLCS | SCLS | SCCS,<numeric_value>

Specifies the capacitance values C1, C2, C3 in the different circuit models for port pair deembedding.  

<Ch>

Channel number  

 

<List_idx>

Current number of a port pair in the list, defined via CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>:DEFine

 

<no>

Number of capacitance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of capacitances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Capacitance C<no> for the specified circuit model.

Range [def. unit]

–1mF to 1 mF [F]. The increment is 1 fF (10–15 F).

*RST value

1 pF (10–12 F)

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the capacitance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:PPA:DEEM:DEF 1,2; :CALC:TRAN:VNET:PPA:DEEM1:PAR:C2? CSSL

Create a list of port pairs and query the default capacitance C2 for the Serial Cs, shunt L circuit model. The response is 1E-012 (1 pF).

CALC:TRAN:VNET:PPA:DEEM:PAR:C2 CSSL, 2.2E-12

Increase the capacitance to 2.2 pF.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>:PARameters:L<no>
STSL | SLST | CSSL | LSSC | LSSL | SLCS | SCLS | SLLS,<numeric_value>

Specifies the inductance values L1, L2, L3 in the different circuit models for port pair deembedding.  

<Ch>

Channel number  

 

<List_idx>

Current number of a port pair in the list, defined via CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>:DEFine

 

<no>

Number of inductance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of inductances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Inductance L<no> for the specified circuit model.

Range [def. unit]

–1H to 1 H [H]. The increment is 1 pH (10–12 H).

*RST value

1 nH (10–9 H)

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the inductance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:PPA:DEEM:DEF 1,2; :CALC:TRAN:VNET:PPA:DEEM:PAR:L1? CSSL

Create a list of port pairs and query the default inductance L1 for the Serial Cs, shunt L circuit model. The response is 1E-009 (1 nH).

CALC:TRAN:VNET:PPA:DEEM:PAR:L1 CSSL, 2.2E-9

Increase the inductance to 2.2 nH.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>:PARameters:R<no>
STSL | STSC | SLST | SCST | CSSL | LSSC | CSSC | LSSL | SLCS | SCLS | SCCS | SLLS,<numeric_value>

Specifies the resistance values R1, R2, R3 in the different circuit models for port pair deembedding.  

<Ch>

Channel number  

 

<List_idx>

Current number of a port pair in the list, defined via CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>:DEFine  

 

<no>

Number of resistance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of resistances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Resistance R<no> for the specified circuit model.

Range [def. unit]

–10 MΩ to 10 MΩ [Ω]. The increment is 1 mΩ (10–3 Ω).

*RST value

0 Ω for all resistances connected in series with an inductance

10 MΩ for all resistances connected in parallel with a capacitance

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the resistance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:PPA:DEEM:DEF 1,2;
CALC:TRAN:VNET:PPA:DEEM:PAR:R1? CSSL; R2? CSSL; R3? CSSL

Create a list of port pairs and query the default resistances for the Serial Cs, shunt L circuit model. The response is 10000000000;10000000000; 0.

CALC:TRAN:VNET:PPA:DEEM:PAR:R3 CSSL, 2.2E+3

Increase the resistance R3 to 2.2 kΩ.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>[:STATe] <Boolean>

Enables or disables the deembedding function for port pairs. It is allowed to change the circuit model and its parameters while deembedding is enabled.

<Ch>

Channel number  

 

<List_idx>

Current number of a port pair in the list, defined via CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>:DEFine  

 

<Boolean>

ON – Deembedding active
OFF – Deembedding inactive

*RST value

OFF [–]

 

SCPI, Command Types

Device-specific, command or query

 

Example:

*RST; :CALC:TRAN:VNET:PPA:DEEM:DEF 1,2,3,4

Define aport pair  configuration with two port pairs.

CALC:TRAN:VNET:PPA:DEEM1:TND CSSL

Select the Serial Cs, shunt L circuit model for the first port pair.

CALC:TRAN:VNET:PPA:DEEM:PAR:R3 CSSL, 2.2E+3; :CALC:TRAN:VNET:PPA:DEEM ON

Increase the resistance R3 for the Serial Cs, shunt L circuit model to 2.2 kΩ and enable deembedding.

...
C
ALC:TRAN:VNET:PPA:DEEM:DEL

Delete the port pair  configuration.

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>:TNDefinition
FIMPort | STSL | STSC | SLST | SCST | CSSL | LSSC | CSSC | LSSL | SLCS | SCLS | SCCS | SLLS

Selects the circuit model for port pair deembedding.

<Ch>

Channel number  

 

<List_idx>

Current number of a port pair in the list, defined via CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding<List_idx>:DEFine

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

CSSL

 

SCPI, Command Types

Device-specific, command or query.  

 

Example:

See CALCulate<Ch>:TRANsform:VNETworks:PPAir:DEEMbedding[:STATe]

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>:DEFine
<first port no>, <second port no>{, <first port no>, <second port no>}

Defines a list of port pairs for port pair embedding. The command can be used repeatedly to extend or overwrite the list.   

<Ch>

Channel number   

 

<List_idx>

List index, denotes the current number of the defined pair of ports within the list. If several port pairs are defined, <List_idx> is the number of the first port pair.  

 

<first port no>,
<second port no>

Two different analyzer ports

Range [def. unit]

1 to 4 (depending on analyzer model) [–]

*RST value

 

SCPI, Command Types

Device-specific, no query

 

Example:

See CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>[:STATe]

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>:DELete

Deletes the previously defined list of port pairs for port pair embedding.  

<Ch>

Channel number   

 

<List_idx>

List index, is available for systematic reasons, but must be always 1 or omitted. The list index in the other ...PPAir:EMBedding... commands denotes the current number of a pair of ports within the list.   

 

*RST value

 

SCPI, Command Types

Device-specific, no query

 

Example:

See CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>[:STATe]  

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>:PARameters:C<no>
STSC | SCST | CSSL | LSSC | CSSC | SLCS | SCLS | SCCS,<numeric_value>

Specifies the capacitance values C1, C2, C3 in the different circuit models for port pair embedding.  

<Ch>

Channel number

 

<List_idx>

Current number of a port pair in the list, defined via CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>:DEFine

 

<no>

Number of capacitance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of capacitances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Capacitance C<no> for the specified circuit model.

Range [def. unit]

–1mF to 1 mF [F]. The increment is 1 fF (10–15 F).

*RST value

1 pF (10–12 F)

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the capacitance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:PPA:EMB:DEF 1,2; :CALC:TRAN:VNET:PPA:EMB:PAR:C2? CSSL

Query the default capacitance C2 for the Serial Cs, shunt L circuit model. The response is 1E-012 (1 pF).

CALC:TRAN:VNET:PPA:EMB:PAR:C2 CSSL, 2.2E-12

Increase the capacitance to 2.2 pF.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>:PARameters:L<no>
STSL | SLST | CSSL | LSSC | LSSL | SLCS | SCLS | SLLS,<numeric_value>

Specifies the inductance values L1, L2, L3 in the different circuit models for port pair embedding.  

<Ch>

Channel number  

 

<List_idx>

Current number of a port pair in the list, defined via CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>:DEFine

 

<no>

Number of inductance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of inductances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Inductance L<no> for the specified circuit model.

Range [def. unit]

–1H to 1 H [H]. The increment is 1 pH (10–12 H).

*RST value

1 nH (10–9 H)

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the inductance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:PPA:EMB:DEF 1,2; :CALC:TRAN:VNET:PPA:EMB:PAR:L1? CSSL

Query the default inductance L1 for the Serial Cs, shunt L circuit model. The response is 1E-009 (1 nH).

CALC:TRAN:VNET:PPA:EMB:PAR:L1 CSSL, 2.2E-9

Increase the inductance to 2.2 nH.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>:PARameters:R<no>
STSL | STSC | SLST | SCST | CSSL | LSSC | CSSC | LSSL | SLCS | SCLS | SCCS | SLLS,<numeric_value>

Specifies the resistance values R1, R2, R3 in the different circuit models for port pair embedding.  

<Ch>

Channel number  

 

<List_idx>

Current number of a port pair in the list, defined via CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>:DEFine

 

<no>

Number of resistance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of resistances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Resistance R<no> for the specified circuit model.

Range [def. unit]

–10 MΩ to 10 MΩ [Ω]. The increment is 1 mΩ (10–3 Ω).

*RST value

0 Ω for all resistances connected in series with an inductance

10 MΩ for all resistances connected in parallel with a capacitance

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the resistance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:PPA:EMB:DEF 1,2
CALC:TRAN:VNET:PPA:EMB:PAR:R1? CSSL; R2? CSSL; R3? CSSL

Query the default resistances for the Serial Cs, shunt L circuit model. The response is 10000000000;10000000000; 0.

CALC:TRAN:VNET:PPA:EMB:PAR:R3 CSSL, 2.2E+3

Increase the resistance R3 to 2.2 kΩ.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>[:STATe] <Boolean>

Enables or disables the embedding function for port pairs. It is allowed to change the circuit model and its parameters while embedding is enabled.

<Ch>

Channel number  

 

<List_idx>

Current number of a port pair in the list, defined via CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>:DEFine

 

<Boolean>

ON – Embedding active
OFF – Embedding inactive

*RST value

OFF [–]

 

SCPI, Command Types

Device-specific, command or query

 

Example:

*RST; :CALC:TRAN:VNET:PPA:EMB:DEF 1,2,3,4

Define a port pair configuration with two port pairs.

CALC:TRAN:VNET:PPA:EMB1:TND CSSL

Select the Serial Cs, shunt L circuit model for port pair no. 1.

CALC:TRAN:VNET:PPA:EMB:PAR:R3 CSSL, 2.2E+3; :CALC:TRAN:VNET:PPA:EMB ON

Increase the resistance R3 for the Serial Cs, shunt L circuit model to 2.2 kΩ and enable embedding.

...
CALC:TRAN:VNET:PPA:EMB:DEL  

Delete the port pair  configuration.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>:TNDefinition
FIMPort | STSL | STSC | SLST | SCST | CSSL | LSSC | CSSC | LSSL | SLCS | SCLS | SCCS | SLLS

Selects the circuit model for port pair embedding.  

<Ch>

Channel number  

 

<List_idx>

Current number of a port pair in the list, defined via CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMBedding<List_idx>:DEFine

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

CSSL

 

SCPI, Command Types

Device-specific, command or query.  

 

Example:

See CALCulate<Ch>:TRANsform:VNETworks:PPAir:EMbedding[:STATe]

 

 


CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>:PARameters:C<no>
CSL | LSC | CSC | SLC | SCL | SCC,<numeric_value>

Specifies the capacitance values C1, C2 in the different circuit models for single ended port deembedding.  

<Ch>

Channel number  

 

<Ph_pt>

Physical port number

 

<no>

Number of capacitance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 2 [–] The number of capacitances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Capacitance C<no> for the specified circuit model.

Range [def. unit]

–1mF to 1 mF [F]. The increment is 1 fF (10–15 F).

*RST value

1 pF (10–12 F)

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the capacitance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:SEND:DEEM:PAR:C2? CSC

Query the default capacitance C2 for the Serial C, shunt C circuit model. The response is 1E-012 (1 pF).

CALC:TRAN:VNET:SEND:DEEM:PAR:C2 CSC, 2.2E-12

Increase the capacitance to 2.2 pF.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>:PARameters:L<no>
CSL | LSC | LSL | SLC | SCL | SLL,<numeric_value>

Specifies the inductance values L1, L2, L3 in the different circuit models for single ended port deembedding.  

<Ch>

Channel number.  

 

<Ph_pt>

Physical port number

 

<no>

Number of inductance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of inductances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Inductance L<no> for the specified circuit model.

Range [def. unit]

–1H to 1 H [H]. The increment is 1 pH (10–12 H).

*RST value

1 nH (10–9 H)

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the inductance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:SEND:DEEM:PAR:L1? SLL

Query the default inductance L1 for the Shunt L, serial L circuit model. The response is 1E-009 (1 nH).

CALC:TRAN:VNET:SEND:DEEM:PAR:L1 SLL, 2.2E-9

Increase the inductance to 2.2 nH.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>:PARameters:R<no>
CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL,<numeric_value>

Specifies the resistance values R1, R2, R3 in the different circuit models for single ended port deembedding.  

<Ch>

Channel number.  

 

<Ph_pt>

Physical port number

 

<no>

Number of resistance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of resistances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Resistance R<no> for the specified circuit model.

Range [def. unit]

–10 MΩ to 10 MΩ [Ω]. The increment is 1 mΩ (10–3 Ω).

*RST value

0 Ω for all resistances connected in series with an inductance

10 MΩ for all resistances connected in parallel with a capacitance

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the resistance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:SEND:DEEM:PAR:R1? CSL; R2? CSL

Query the default resistances for the Serial C, shunt L circuit model. The response is 10000000; 0.

CALC:TRAN:VNET:SEND:DEEM:PAR:R2 CSL, 2.2E+3

Increase the resistance R2 to 2.2 kΩ.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>[:STATe] <Boolean>

Enables or disables the deembedding function for single ended ports. It is allowed to change the circuit model and its parameters while deembedding is enabled.

<Ch>

Channel number.  

 

<Ph_pt>

Physical port number

 

<Boolean>

ON – Deembedding active
OFF – Deembedding inactive

*RST value

OFF [–]

 

SCPI, Command Types

Device-specific, command or query

 

Example:

CALC:TRAN:VNET:SEND:DEEM:TND CSL

Select the Serial C, shunt L circuit model for deembedding.

CALC:TRAN:VNET:SEND:DEEM:PAR:R2 CSL, 2.2E+3; :CALC:TRAN:VNET:SEND:DEEM ON

Increase the resistance R2 for the Serial C, shunt L circuit model to 2.2 kΩ and enable deembedding.

 

 


CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding<Ph_pt>:TNDefinition
FIMPort | CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL

Selects the circuit model for single ended port deembedding.  

<Ch>

Channel number.  

 

<Ph_pt>

Physical port number

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

CSL

 

SCPI, Command Types

Device-specific, command or query.  

 

Example:

See CALCulate<Ch>:TRANsform:VNETworks:SENDed:DEEMbedding[:STATe]

 

 


CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<Ph_pt>:PARameters:C<no>
CSL | LSC | CSC | SLC | SCL | SCC,<numeric_value>

Specifies the capacitance values C1, C2 in the different circuit models for single ended port embedding.  

<Ch>

Channel number.  

 

<Ph_pt>

Physical port number

 

<no>

Number of capacitance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 2 [–] The number of capacitances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Capacitance C<no> for the specified circuit model.

Range [def. unit]

–1mF to 1 mF [F]. The increment is 1 fF (10–15 F).

*RST value

1 pF (10–12 F)

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the capacitance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:SEND:EMB:PAR:C2? CSC

Query the default capacitance C2 for the Serial C, shunt C circuit model. The response is 1E-012 (1 pF).

CALC:TRAN:VNET:SEND:EMB:PAR:C2 CSC, 2.2E-12

Increase the capacitance to 2.2 pF.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<Ph_pt>:PARameters:L<no>
CSL | LSC | LSL | SLC | SCL | SLL,<numeric_value>

Specifies the inductance values L1, L2, L3 in the different circuit models for single ended port embedding.  

<Ch>

Channel number.  

 

<Ph_pt>

Physical port number

 

<no>

Number of inductance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of inductances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Inductance L<no> for the specified circuit model.

Range [def. unit]

–1H to 1 H [H]. The increment is 1 pH (10–12 H).

*RST value

1 nH (10–9 H)

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the inductance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:SEND:EMB:PAR:L1? SLL

Query the default inductance L1 for the Shunt L, serial L circuit model. The response is 1E-009 (1 nH).

CALC:TRAN:VNET:SEND:EMB:PAR:L1 SLL, 2.2E-9

Increase the inductance to 2.2 nH.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<Ph_pt>:PARameters:R<no>
CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL,<numeric_value>

Specifies the resistance values R1, R2, R3 in the different circuit models for single ended port embedding.  

<Ch>

Channel number.  

 

<Ph_pt>

Physical port number

 

<no>

Number of resistance in circuit model. If unspecified the numeric suffix is set to 1.

Range [def. unit]

1 to 3 [–] The number of resistances depends on the selected circuit model.

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

– (see command type description below)

 

<numeric_value>

Resistance R<no> for the specified circuit model.

Range [def. unit]

–10 MΩ to 10 MΩ [Ω]. The increment is 1 mΩ (10–3 Ω).

*RST value

0 Ω for all resistances connected in series with an inductance

10 MΩ for all resistances connected in parallel with a capacitance

 

SCPI, Command Types

Device-specific, command or query. In the query form, the <numeric_value> parameter must be omitted. The command returns the resistance value for the specified circuit model.

 

Example:

*RST; :CALC:TRAN:VNET:SEND:EMB:PAR:R1? CSL; R2? CSL

Query the default resistances for the Serial C, shunt L circuit model. The response is 10000000; 0.

CALC:TRAN:VNET:SEND:EMB:PAR:R2 CSL, 2.2E+3

Increase the resistance R2 to 2.2 kΩ.  

 

 


CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<Ph_pt>[:STATe] <Boolean>

Enables or disables the embedding function for single ended ports. It is allowed to change the circuit model and its parameters while embedding is enabled.

<Ch>

Channel number.  

 

<Ph_pt>

Physical port number

 

<Boolean>

ON – Embedding active
OFF – Embedding inactive

*RST value

OFF [–]

 

SCPI, Command Types

Device-specific, command or query

 

Example:

CALC:TRAN:VNET:SEND:EMB:TND CSL

Select the Serial C, shunt L circuit model for embedding.

CALC:TRAN:VNET:SEND:EMB:PAR:R2 CSL, 2.2E+3; :CALC:TRAN:VNET:SEND:EMB ON

Increase the resistance R2 for the Serial C, shunt L circuit model to 2.2 kΩ and enable embedding.

 

 


CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<Ph_pt>:TNDefinition
FIMPort | CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL

Selects the circuit model for single ended port embedding.  

<Ch>

Channel number.  

 

<Ph_pt>

Physical port number

 

Parameters

Possible circuit models (character data); see circuit models.

*RST value

CSL

 

SCPI, Command Types

Device-specific, command or query.  

 

Example: