This subsystem controls the horizontal line used to mark and retrieve response values (display line).
Command tree: CALCulate<Chn>
:DLINe
[:STATe]
Defines the position (response value) of the horizontal line.
<Chn>
Channel number used to identify the active trace. If unspecified the numeric suffix is set to 1.
Parameters
See list of parameters below.
Range [def. unit]
The data is transferred in the data format defined via FORMat[:DATA]. The unit is the default unit of the measured parameter; see CALCulate<Ch>:PARameter:SDEFine.
*RST value
–
SCPI, Command Types
Device-specific, command or query.
Example:
*RST; :CALC:DLIN 10
Define the position of the horizontal line in the default dB Mag diagram at +10 dBm.
CALC:DLIN:STAT ON
Display the defined horizontal line.
Switches the horizontal line on or off.
Channel number used to identify the active trace.
<Boolean>
ON | OFF - Horizontal line on or off.
OFF
See CALC:DLIN.