CALCulate<Chn>:FORMat...    Related Topics

This subsystem determines a post-processing of the measured data in order to obtain various display formats.

 

Command tree: CALCulate<Chn>

:FORMat

:WQUType


CALCulate<Chn>:FORMat
MLINear | MLOGarithmic | PHASe | UPHase | POLar | SMITh | ISMith | GDELay | REAL | IMAGinary | SWR | COMPlex | MAGNitude

Defines how the measured result at any sweep point is post-processed and presented in the graphical display.

Note: The analyzer allows arbitrary combinations of display formats and measured quantities; see Trace - Format and CALCulate<Chn>:PARameter commands. Nevertheless, it is advisable to check which display formats are generally appropriate for an analysis of a particular measured quantity; see Measured Quantities and Display Formats.

<Chn>

Channel number used to identify the active trace. If unspecified the numeric suffix is set to 1.

 

Parameters

See list of parameters below.

*RST value

MLOGarithmic

 

SCPI, Command Types

Confirmed, command or query

 

Example:

CALC4:PAR:SDEF 'Ch4Tr1', 'S11'

Create channel 4 and a trace named Ch4Tr1 to measure the input reflection coefficient S11. The trace becomes the active trace in channel 4.

CALC4:FORM MLIN; :DISP:WIND:TRAC2:FEED 'CH4TR1'

Calculate the magnitude of S11 and display it in a linearly scaled Cartesian diagram assigning the trace number 2.

 

 

Assume that the result at a sweep point is given by the complex quantity z = x + jy. The meaning of the parameters is as follows (see also table in CALCulate<Chn>:MARKer<Mk>:FORMat description):

MLINear

Calculate the magnitude of z, to be displayed in a Cartesian diagram with a linear scale

MLOGarithmic

Calculate the magnitude of z, displayed in a Cartesian diagram with a logarithmic scale

PHASe

Phase of z, displayed in a Cartesian diagram with a linear vertical axis

UPHase

Unwrapped phase of z, displayed in a Cartesian diagram with a linear vertical axis

POLar

Magnitude and phase, displayed in a polar diagram

SMITh

Magnitude and phase, displayed in a Smith chart

ISMith

Magnitude and phase, displayed in an inverted Smith chart

GDELay

Group delay, displayed in a Cartesian diagram

REAL

Real part (x), displayed in a Cartesian diagram

IMAGinary

Imaginary part (y), displayed in a Cartesian diagram

SWR

Standing Wave Ratio (SWR), displayed in a Cartesian diagram

COMPlex

x, y, displayed in a polar diagram

MAGNitude

Magnitude (sqrt(x2 + y2)), displayed in a Cartesian diagram with a logarithmic scale

 


CALCulate<Chn>:FORMat:WQUType POWer | VOLTage

Selects the physical unit of the displayed trace.

<Chn>

Channel number used to identify the active trace

 

Parameters

Power or voltage units

*RST value

POWer

 

SCPI, Command Types

Confirmed, command or query

 

Example:

CALC4:PAR:SDEF 'Ch4Tr1', 'b1'

Create channel 4 and a trace named Ch4Tr1 to measure the wave quantity b1. The trace becomes the active trace in channel 4.

CALC4:FORM:WQUT VOLT

Select voltage units for the created trace (identified by the suffix 4).