[SENSe<Ch>:]POWer...    Related Topics

This subsystem controls the power at the input ports of the analyzer.

Command tree: [SENSe<Ch>:]POWer

:IFGain

:MEASure

:REFerence


 


[SENSe<Ch>:]POWer:IFGain<Pt>:MEASure
AUTO | LNOise | LDIStortion

Selects the IF gain in the measurement channel (b-waves).

<Ch>

Channel number.

 

<Pt>

Test port number of the analyzer.

 

Parameters

AUTO – Adaptive Gain Control according to the RF input level
LNOise – Fixed, large IF gain, for low input levels  
LDIStortion – Fixed, small IF gain, for high input levels
 

*RST value

AUTO

 

SCPI, Command Types

Device-specific, command or query

 

Example:

POWer:IFG2:MEAS LDIS; POWer:IFG2:REF LNO

Optimize the measurement channel b for small input levels, the reference channel a for high input levels.

 

  


[SENSe<Ch>:]POWer:IFGain<Pt>:REFerence
AUTO | LNOise | LDIStortion

Selects the IF gain in the reference channel (a-waves). This setting is not available on R&S ZVB4, R&S ZVB8, R&S ZVB20, and R&S ZVT20 analyzers.

<Ch>

Channel number.

 

<Pt>

Test port number of the analyzer.

 

Parameters

AUTO – Adaptive Gain Control according to the RF input level
LNOise – Fixed, small IF gain, for high input levels  
LDIStortion – Fixed, large IF gain, for low input levels
 

*RST value

AUTO

 

SCPI, Command Types

Device-specific, command or query

 

Example:

See [SENSe<Ch>:]POWer:IFGain<Pt>:MEASure