SOURce<Ch>:POWer<Pt>:CONVerter...    Related Topics

This subsystem provides general frequency converter settings.

   

Command tree: [SENSe<Ch>:]POWer<Pt>:CONVerter...

:OFFSet

:TRANsfer

:AMODel

:ATTenuator

:DESCription

:ELECtronic

:LIMit

:MATTenuation

:REDuction

:MECHanical

:ATTenuation

:OFFSet

:SLOPe


SOURce<Ch>:POWer<Pt>:CONVerter:OFFSet
<offset>, ONLY | CPADd

Defines a source power or a power offset relative to the channel power (SOURce<Ch>:POWer<Pt>[:LEVel][:IMMediate][:AMPlitude]) for a frequency converter port <Pt>. The command is available for converters with electronic attenuators (R&S ZVA-ZxxxE). An additional Cal Power Offset can be defined via SOURce<Ch>:POWer<Pt>:CORRection:LEVel:OFFSet.

<Ch>

Channel number

 

<Pt>

Converter port number

 

<offset>

Port-specific power offset

Range [def. unit]

–300 dB to +300 dB [dB]. UP and DOWN increment/decrement the phase in 1-dB steps.

*RST value

0 dB

 

Offset parameters

ONLY – only the port-specific power is used; the channel power is ignored.

CPADd – the port-specific power is added as an offset to the channel power.

*RST value

CPADd

 

SCPI, Command Types

Device-specific, command or query

 

Example:

See SOURce<Ch>:FREQuency<Pt>:CONVersion:ARBitrary:CFRequency

 

 


SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:AMODel <Boolean>

Enables or disables the advanced power transfer model for the converter port <Pt>.

This command is overwritten by the general power transfer model settings; see [SENSe<Ch>:]CONVerter:AMODel.

<Ch>

Channel number.

 

<Pt>

Converter port number

 

<Boolean>

ON | OFF - Enables or disables the advanced model.

*RST value

OFF

 

SCPI, Command Types

Device-specific, command or query

 

Example:

SOUR:POW2:CONV:TRAN:AMOD ON; DESC LAPP

Enable the advanced power transfer model at the frequency converter port 2 and select a linear approximation.

SOUR:POW2:CONV:TRAN:SLOP 1; OFFS 0

Define the slope and offset for the linear model with0 dB waveguide attenuators setting.

SOUR:POW2:CONV:TRAN:ATT MECH
SOUR:POW2:CONV:TRAN:MECH:ATT 3

Select a converter type with a mechanical waveguide attenuator and define a 3-dB attenuation

Alternatively....

SOUR:POW2:CONV:TRAN:ATT ELEC
SOUR:POW2:CONV:TRAN:ELEC:LIM -5; RED 50; MATT 20

Select a converter type with electronic attenuators and define a power threshold of -5 dBm, 50 %  power reduction due to the waveguide attenuator, and a maximum attenuation of 20 dB.

 

  


SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:ATTenuator MECHanical | ELECtronic

Selects the waveguide attenuator type of the converter port <Pt>.

<Ch>

Channel number.

 

<Pt>

Converter port number

 

Parameters

MECHanical - Adjustment screw (most converter models R&S ZVA-Zxxx)  
ELECtronic - Electronic attenuator (converters R&S ZVA.ZxxxE)

*RST value

MECH

 

SCPI, Command Types

Device-specific, command or query

 

Example:

See SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:AMODel

 

 


SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:DESCription DSET | LAPProx

Selects the type of advanced power transfer model for the converter port <Pt>.

This command is overwritten by the general power transfer model settings; see [SENSe<Ch>:]CONVerter:DESCription.

<Ch>

Channel number.

 

<Pt>

Converter port number

 

Parameters

DSET - Data set (for future extensions)
LAPProx - Linear approximation  

*RST value

n/a (a *RST does not affect the general frequency converter settings)

 

SCPI, Command Types

Device-specific, command or query

 

Example:

See SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:AMODel

 

 


SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:ELECtronic:LIMit <threshold>

Specifies the converter output power threshold for a converter with electronic attenuators at port <Pt>.

<Ch>

Channel number.

 

<Pt>

Converter port number

 

<threshold>

Converter output power below which the waveguide attenuator is used  

Range [def. unit]

–300 dBm to +200 dBm [dBm], the increment (UP | DOWN) is 0.1 dB

*RST value

–5 dBm

 

SCPI, Command Types

Device-specific, command or query

 

Example:

See SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:AMODel

 

 


SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:ELECtronic:MATTenuation <attenuation>

Specifies the maximum attenuation for a converter with electronic attenuators at port <Pt>.  

<Ch>

Channel number.

 

<Pt>

Converter port number

 

<attenuation>

Maximum attenuation due to the waveguide attenuator

Range [def. unit]

0 dB to INF [dB], the increment (UP | DOWN) is 0.1 dB

*RST value

20 dB

 

SCPI, Command Types

Device-specific, command or query

 

Example:

See SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:AMODel

 

 


SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:ELECtronic:REDuction <percentage>

Specifies the percentage of power reduction for a converter with electronic attenuators at port <Pt>.  

<Ch>

Channel number.

 

<Pt>

Converter port number

 

<percentage>

Percentage of power reduction due to the waveguide attenuator  

Range [def. unit]

1 % to 100 % [%], the increment (UP | DOWN) is 1 %

*RST value

50 %

 

SCPI, Command Types

Device-specific, command or query

 

Example:

See SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:AMODel

 

 


SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:MECHanical:ATTenuation <attenuation>

Specifies the attenuation factor for a converter with mechanical adjustment screw at port <Pt>.  

<Ch>

Channel number.

 

<Pt>

Converter port number

 

<attenuation>

Attenuation due to the adjustment screw  

Range [def. unit]

0 dB to INF  [dB], the increment (UP | DOWN) is 0.1 dB

*RST value

8 dB

 

SCPI, Command Types

Device-specific, command or query

 

Example:

See SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:AMODel

 

 


SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:OFFSet <offset>

Specifies the offset factor for the linear power transfer model with 0 dB waveguide attenuation.

<Ch>

Channel number.

 

<Pt>

Converter port number

 

Parameters

Offset in the linear power transfer model

Range [def. unit]

–300 dB to +300 dB [dB], the increment (UP | DOWN) is 1 dB

*RST value

0 dB

 

SCPI, Command Types

Device-specific, command or query

 

Example:

See SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:AMODel

 

 


SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:SLOPe <slope>

Specifies the slope for the linear power transfer model with 0 dB waveguide attenuation.

<Ch>

Channel number.

 

<Pt>

Converter port number

 

<slope>

Slope in the linear power transfer model  

Range [def. unit]

1 to 1000000 [1], the increment (UP | DOWN) is 1

*RST value

1

 

SCPI, Command Types

Device-specific, command or query

 

Example:

See SOURce<Ch>:POWer<Pt>:CONVerter:TRANsfer:AMODel