Noise Figure Measurement    Related Topics  

The Noise Figure Meas submenu controls the measurement of the noise figure of a DUT. This measurement requires option R&S ZVAB-K30.

Noise Figure Measurements

The noise figure of a device is the ratio of the output signal-to-noise ratio (SNR) to the input SNR, provided that a thermal noise signal is fed to the input. It is a measure of the degradation of the SNR caused by the device.

The purpose of the Noise Figure Meas is to assess the noise figure of a DUT which operates in its linear range. A typical example is an amplifier which is measured in transmission. The method of measurement offers several advantages:

Noise figure measurements can be performed irrespective of the sweep type: A frequency sweep yields the noise figure as a function of the frequency, a power sweep yields the power-dependent noise figure at fixed frequency. See also Basic Test Setup and Noise Figure Calculation.  

Refer to application note 1EZ61 for a detailed discussion of the noise figure measurement including many examples.

Performing a noise figure measurement


Basic Test Setup and Noise Figure Calculation

The schematic test setup for a noise figure measurement is shown below.

The NWA source provides a superposition of a CW signal and a noise signal. The latter generally consists of the thermal noise NT, 1 and a possible additional contribution ΔN:

PNoise, 1 = NT, 1 + ΔN

The definition of the noise figure FD is based on a pure thermal noise signal (ΔN = 0):

FD = (PSignal, 1 / NT, 1) / (PSignal, 2 / PNoise, 2) = PNoise, 2 / (NT, 1 * GD) where GD = PSignal, 2 / PSignal, 1

ΔN is supposed to be amplified like the CW signal. Hence, the noise signal at the output of the DUT equals to:

PNoise, 2 = GD * (FD * NT + ΔN)

According to the equations above, it is possible to calculate both the gain GD and the noise figure FD of the DUT, if the four powers PSignal, 2, PSignal, 1, PNoise, 1, PNoise, 2 can be measured with sufficient accuracy. The measurement requires two steps:

Conditions for the measurement

The Noise Figure Setup Guide helps you perform all the necessary steps consistently and in the right order.

Extensions

Performing a noise figure measurement


Noise Figure Setup Guide

The Noise Figure Setup Guide gives access to all necessary steps of a noise figure measurement. The dialog is similar to a wizard, however, everything is arranged in a single dialog. The panels numbered 1. to 7. suggest a meaningful order for the different steps. In some cases, it is possible to change the order or skip a step entirely (e.g. a calibration may be skipped if a valid calibration for the test setup is already available).

The buttons in the different panes open dialogs or wizards; all of them are also accessible from the menus of the network analyzer. In general, it is advantageous though to open the dialogs from the setup guide, because this will adjust some of the settings to the requirements of the noise figure measurement.  

The setup guide provides the following panels/steps:

1. DUT Port Connection defines the basic test setup, i.e. the input and output ports of the DUT/NWA. With option R&S ZVA-B16, Direct Generator and Receiver Access, it is possible to measure the noise figure at a single port (<DUT Input> = <DUT Output>); see section Basic Test Setup and Noise Figure Calculation.

 

2. Source Power Calibration gives access to the Source Power Cal dialog for the previously selected DUT Input port and base power. The base power setting overwrites the Channel – Power Bandwidth Average Power setting as soon as the calibration dialog is opened.   

 

3. Receiver Power Calibration gives access to the Receiver Power Cal dialog for the previously selected DUT Output port. If a preamplifier is connected between the DUT and the receive port of the analyzer, or if the output signal of the DUT is fed to the MEAS IN connector of option R&S ZVA-B16, Direct Generator and Receiver Access, in order to bypass the coupler and improve the receiver noise figure, the check box should be selected. This activates the input fields for base power and source step attenuator settings (if available), which you may have to adjust to the test setup in the receiver path in order to avoid excess receiver input powers (receiver compression; see data sheet). An active check box has no additional effects.
The base power setting overwrites the Channel – Power Bandwidth Average Power setting as soon as the calibration dialog is opened. The step attenuator setting overwrites the source attenuator settings for the DUT Input port in the Channel – Power Bandwidth Average Step Attenuators dialog.

 

4. Define Noise Figure opens the Define Noise Figure Measurement dialog, providing basic settings for the noise figure measurement.  

 

5. Noise Figure Calibration opens the Noise Figure Calibration dialog to assess and compensate the internal noise at the NWA receiver and source.  

 

6. S-Parameter Calibration opens the Calibration wizard for the previously selected input and output ports and the Cal Method. If this step is skipped, the previous power and noise figure calibrations effectively ensure a transmission normalization; the noise figure trace line contains no calibration label. With an additional One Path Two Port or Full Two Port calibration, the system error correction of the gain GD (and thus the noise figure measurement in total) is improved; the trace lines for the calibrated ports indicate Cal. A Full Two Port calibration requires more calibration measurements but provides best accuracy; see section Calibration Types.
Note:
An S-parameter calibration is required for S-parameter measurements in parallel to the noise figure measurement. The large IF bandwidth required for noise figure measurements limits the accuracy of the calibration and the subsequent S-parameter measurements; the AVG detector can improve the results.

 

7. Configure Traces provides a selection of traces which are commonly of interest for the noise figure measurement. The selected traces are created and displayed when the Replace or Add buttons are clicked. The noise figure trace (NF21) is smoothed automatically (label SMO in the trace line). The large IF bandwidth limits the accuracy of the S-parameter measurement; the AVG detector can improve the results. Use the Trace – Measure menu if you wish to modify the trace settings.
To avoid misleading results, S-parameter selection is enabled only after an S-Parameter Calibration has been performed.   


Define Noise Figure Meas...

Provides basic parameters for the noise figure measurement.  

Sequential versus Simultaneous measurement

During the noise figure measurement stage, the analyzer must determine the powers of the CW signal and of the noise signal at the output of the DUT (PSignal, 2, PNoise, 2). This may be done in a single or in two measurement stages.

Remote control:

[SENSe<Ch>:]SWEep:DETector:TIME
[SENSe<Ch>:]NFIGure:SEQuential


Noise Figure Cal...

Opens a dialog to assess and compensate the internal noise at the NWA receiver and source. The noise figure calibration should be performed after regular power calibration using an external power sensor following the instructions in section Power Calibration. It is required for various reasons (see also Basic Test Setup and Noise Figure Calculation):

The necessary calibration steps are automatically performed with the channel settings of the active noise figure measurement.

Noise Figure Calibration procedure

The noise figure calibration is performed in two steps. A third step is added if an external attenuator is used.

1. Receiver noise calibration of the analyzer's receive port. The receive port (assigned to the DUT output) is terminated with a 50 Ω match. The connecting cable for the DUT should also be used to connect the match. If a pre-amplifier (preferably, the low-noise preamplifier R&S ZVAX-B203) is used for the measurement, it is also part of the calibration test setup. The dot in the figure below denotes the position of the DUT output in the measurement test setup.

The noise figure of the NWA receiver (including the connecting cable and a possible pre-amplifier) limits the accuracy of the noise figure measurement as a whole. A pre-amplifier with a higher gain can reduce the receiver noise figure.

2. Source noise calibration of the analyzer's source port. The DUT is replaced by a through connection; its attenuation should be as small as possible. If a pre-amplifier is used for the measurement, it is also part of the calibration test setup. The two dots in the figure below denote the position of the DUT input and output in the measurement test setup.  

3. Attenuator calibration, if the measurement is performed with an external attenuator. The external attenuator and a possible pre-amplifier are part of the calibration test setup. The DUT is replaced by a through connection. The two dots in the figure below denote the position of the DUT input and output in the measurement test setup.  

Remote control:

SENSe1:POWer:IFGain<pt>:MEASure
[SENSe<Ch>:]CORRection:COLLect:NFIGure:STARt

[SENSe<Ch>:]CORRection:COLLect:NFIGure[:ACQuire]

[SENSe<Ch>:]CORRection:COLLect:NFIGure:SAVE

[SENSe<Ch>:]CORRection:COLLect:NFIGure:END


Correction Off

Activates or deactivates the Noise Figure Calibration in the active channel. Correction Off is available only after a calibration has been performed and applied to the active channel.

With deactivated correction, the analyzer calculates an approximate noise figure, neglecting the influence of internal noise, attenuators etc. This can introduce large deviations between the uncorrected and the corrected traces.  

Remote control:

[SENSe<Ch>:]CORRection:NFIGure[:STATe]