Alternating Sweeps    Related Topics

Activates or deactivates the operating mode where the analyzer performs only one partial measurement per sweep.

The Alternating Sweeps mode is also valid during a system error correction.

Use the Alternating Sweeps mode to increase the accuracy of measurements on DUTs with long level settling times (e.g. quartzes, SAW filters). To measure DUTs with short settling times or obtain a trace from the beginning of the sweep, deactivate the Alternating Sweeps mode.

In triggered measurements, Alternate has no effect if the triggered measurement sequence is identical to a single sweep point. The following table shows how the analyzer performs a sweep comprising m sweep points, assuming that each of them requires n partial measurements.

Triggered Meas. Sequence

Alternate On

Alternate Off

Sweep

Trigger event starts n partial sweeps over all sweep points. For a 2-port transmission measurement comprising two partial measurements (S12 and S21):

Trigger event starts m complete measurements at consecutive sweep points. For a 2-port transmission measurement comprising two partial measurements (S12 and S21):

Sweep Segment

Trigger event starts n partial sweeps over the next segment.

Trigger event starts complete measurements at all consecutive sweep points in the segment.  

Point

All partial measurements of each sweep point are carried out one after another.

Same as Alternate On: All partial measurements of each sweep point are carried out one after another.

Partial Measurement

Each partial measurement is carried out for all sweep points.
Source port specific trigger delays are supported.

All partial measurements of each sweep point are carried out one after another
Source port specific trigger delays are supported.

Remote control:

[SENSe<Ch>:]COUPle ALL | NONE